Example of Performing the Bradford Protein Assay Using NanoQ Series

In this application note, the applicability of the NanoQ Plus microvolume spectrophotometer to the Bradford Protein Assay was evaluated. After the color development reaction was carried out using the Pierce™ Bradford Plus Protein Assay Kit, based on Coomassie Brilliant Blue G-250 dye, together with Dilution-Free™ BSA Protein Standards, the samples were pipetted onto the NanoQ Plus pedestal and measured in Bradford Assay mode (595 nm). Measurement of BSA standard solutions ranging from 125 to 2,000 µg/mL showed that the absorbance at 595 nm increased with increasing concentration, and the second-order polynomial calibration curve generated from the measured values yielded a coefficient of determination (R²) of 0.9987. Back-calculation of the concentration of each standard solution using the calibration curve gave relative errors (%RE) ranging from −5.20% to 5.32%. These results confirm that the NanoQ Plus can be used to measure the Bradford Protein Assay color development reaction and to generate a calibration curve.
Overview
The Bradford Protein Assay is a representative colorimetric method that quantifies protein concentration by measuring the change in absorbance that occurs when Coomassie Brilliant Blue G-250 dye binds to protein. The dye develops a blue color upon binding to protein, and the protein concentration of a sample can be quantified by measuring the absorbance of the resulting complex.
NanoQ Plus is a microvolume spectrophotometer widely used for measuring the concentration and purity of nucleic acid (DNA, RNA) and protein samples. NanoQ Plus supports microvolume absorbance measurement using a pedestal and can be used for a variety of absorbance-based assays. This application note introduces an example of performing the Bradford Protein Assay using NanoQ Plus.

Materials and Methods
Reagents and Equipment
This test was performed using the NanoQ Plus microvolume spectrophotometer. Protein quantification was carried out using the Pierce™ Bradford Plus Protein Assay Kit (Thermo Fisher Scientific, Cat. No. A55866), and the Pierce™ Dilution-Free™ BSA Protein Standards (REF 1863881) included in the kit were used as the standard material.
Test Method
A calibration curve was constructed using the 125, 250, 500, 750, 1,000, 1,500, and 2,000 µg/mL standard solutions included in the Pierce™ Dilution-Free™ BSA Protein Standards.
Immediately before use, the Bradford Plus Protein Assay Reagent bottle was gently inverted several times to ensure thorough mixing and then equilibrated to room temperature.
The blank was prepared using 5 µL of distilled water. After adding 250 µL of Bradford Plus Protein Assay Reagent to each standard solution and the blank, the mixtures were vortexed for approximately 30 seconds and allowed to react at room temperature for 10 minutes to complete color development.
Reverse pipetting is a technique in which excess liquid remains inside the pipette tip while the set volume is dispensed. Using reverse pipetting, 2 µL of the reacted sample was applied onto the NanoQ Plus pedestal and measured using the Bradford Assay mode. A polynomial calibration curve was constructed using the standard solution concentrations and the measured absorbance values.
Results
The Bradford Protein Assay using the Pierce™ Dilution-Free™ BSA Protein Standards (125–2,000 µg/mL) generated a polynomial calibration curve in the Bradford Assay mode of NanoQ Plus. The absorbance at 595 nm increased as the concentration of the BSA standard solution increased, and the calibration curve’s R² was 0.9987.
Back-calculating the concentration using the absorbance value of each standard solution, the %RE was confirmed to range from −5.20% to 5.32%.
Table 1. Absorbance and back-calculation results by concentration of BSA standard solutions
| Conc. (µg/mL) | Abs. 595 nm | Back-calculated Conc. (µg/mL) | %RE |
|---|---|---|---|
| 125 | 0.12 | 120.85 | −3.32 |
| 250 | 0.28 | 263.30 | 5.32 |
| 500 | 0.54 | 512.13 | 2.43 |
| 750 | 0.73 | 711.00 | −5.20 |
| 1,000 | 1.00 | 1,027.82 | 2.78 |
| 1,500 | 1.33 | 1,503.50 | 0.23 |
| 2,000 | 1.56 | 1,956.21 | −2.19 |
Conclusion
In this test, the Bradford Protein Assay was performed using the Pierce™ Bradford Plus Protein Assay Kit and NanoQ Plus. A polynomial calibration curve was constructed using the Bradford Assay mode of NanoQ Plus, confirming that NanoQ Plus can be applied to the Bradford Protein Assay.
References (Sources)
- Bradford, M. M. (1976). A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Analytical Biochemistry, 72, 248–254.
- Thermo Fisher Scientific. Pierce™ Bradford Plus Protein Assay Kit with Dilution-Free™ BSA Protein Standards, Multichannel Pipette Compatible, User Guide, Cat. No. A55866. Thermo Fisher Scientific.