Example of Bradford Protein Assay Using the POP UV-Vis Spectrophotometer

POP_Example of Bradford Protein Assay using the POP UV-Vis spectrophotometer

This application note evaluates the applicability of the POP UV-Vis spectrophotometer to the Bradford Protein Assay. Color development was performed using Pierce™ Plus Protein Assay Reagent and Dilution-Free™ BSA Protein Standards. The samples were then transferred to semi-micro cuvettes, and absorbance was measured at 595 nm.
In the Standard Assay range of 125–2,000 µg/mL, the increase in absorbance gradually became less pronounced as the concentration increased, indicating a nonlinear response. A second-order polynomial regression was therefore applied. The coefficient of determination (R²) was 0.9996, and the relative error (%RE) was 2.43–3.56%. In the Micro Assay range of 1–25 µg/mL, a linear relationship was observed between concentration and absorbance, and linear regression gave an R² of 0.9981. These results confirm that the POP UV-Vis spectrophotometer can be applied to protein quantification using the Bradford Protein Assay over both the Standard Assay and Micro Assay concentration ranges.

Introduction

The Bradford Protein Assay is a widely used colorimetric method for protein quantification based on the change in absorbance that occurs when Coomassie Brilliant Blue G-250 dye binds to proteins. Upon binding to protein, the dye develops a blue color, and the protein concentration of a sample can be quantified by measuring the absorbance of the resulting complex at 595 nm. Because the Bradford Assay is rapid and simple to perform, it is widely used as a standard method for protein quantification in life science, biochemistry, and molecular biology.

In this application note, a standard curve was generated using BSA (bovine serum albumin) standards and Pierce™ Bradford Plus Protein Assay Reagent, and the applicability of the Bradford Protein Assay using the POP UV-Vis spectrophotometer and semi-micro cuvettes was evaluated. Standard curves were also generated for two concentration ranges, Standard Assay (125–2,000 µg/mL) and Micro Assay (1.5625–25 µg/mL), to evaluate the applicability of each concentration range.

Figure 1
Figure 1. POP UV-Vis Spectrophotometer from K LAB

Materials and Methods

Reagents and Instrumentation

The assay was performed using the POP UV-Vis spectrophotometer. Pierce™ Bradford Protein Assay Reagent (Thermo Fisher Scientific, Cat. No. A55866) was used for protein quantification, and Pierce™ Dilution-Free™ BSA Protein Standards were used as the standard material. Absorbance measurements were performed using semi-micro cuvettes (Ratiolab, Cat. No. 2810100).

Experimental Procedure

BSA standards were prepared separately for two concentration ranges: Standard Assay (125–2,000 µg/mL) and Micro Assay (1.5625–25 µg/mL). For the Micro Assay, a 25 µg/mL BSA standard was serially diluted twofold with distilled water to prepare five concentrations (1.5625, 3.125, 6.25, 12.5, and 25 µg/mL). The dispensing conditions for the standards and Bradford Plus Protein Assay Reagent for the two assays are shown in Table 1.

Table 1. Dispensing Conditions for BSA Standards and Bradford Reagent by Concentration Range

BSA Concentration Range(µg/mL) BSA Standard Dispensing Volume (µL) Bradford Reagent Dispensing Volume (µL)
125–2,000 20 1,000
1.5625–25 550 550

Immediately before use, the Bradford Plus Protein Assay Reagent bottle was gently inverted several times to mix the reagent thoroughly, and the reagent was equilibrated to room temperature. Each standard was mixed with Bradford Plus Protein Assay Reagent, vortexed for approximately 30 seconds, and incubated at room temperature for 10 minutes to complete color development.

After the reaction was complete, approximately 1,000 µL of each sample was transferred to a semi-micro cuvette, and absorbance at 595 nm was measured in Photometric mode using the POP UV-Vis spectrophotometer. The blank was prepared by mixing distilled water and Bradford Plus Protein Assay Reagent at the same ratio used for each assay.

Figure 2
Figure 2. Bradford Standard Assay Samples (125–2,000 µg/mL) Dispensed into Semi-Micro Cuvettes

Figure 3
Figure 3. Bradford Micro Assay Samples (1.5625–25 µg/mL) Dispensed into Semi-Micro Cuvettes

Results

The Bradford Protein Assay was performed using BSA standards over two concentration ranges, 125–2,000 µg/mL and 1.5625–25 µg/mL. Standard curves were generated from the measured absorbance values, and the back-calculated concentration and relative error (%RE) were calculated to evaluate the applicability of each concentration range.

In the 125–2,000 µg/mL concentration range, the increase in absorbance gradually became less pronounced as the concentration increased. A polynomial regression was therefore applied to generate the standard curve. The standard curve results are shown in Table 2 and Figure 4.

Table 2. Standard Curve Results for the 125–2,000 µg/mL Concentration Range

Input Concentration (µg/mL) Abs. 595 nm Back-calculated Conc.(µg/mL) %RE
125 0.110 123.12 −1.51
250 0.276 258.91 3.56
500 0.541 497.54 −0.49
750 0.768 731.76 −2.43
1,000 0.992 1,005.89 0.59
1,500 1.294 1,527.18 1.81
2,000 1.422 1,964.41 −1.78

Figure 4
Figure 4. Bradford Protein Assay Standard Curve for the 125–2,000 µg/mL Concentration Range Using the POP UV-Vis Spectrophotometer A second-order polynomial regression was applied to generate the standard curve, and the coefficient of determination (R²) was 0.9996.

In the 1.5625–25 µg/mL concentration range, a linear relationship was observed between absorbance and concentration. A linear regression was therefore applied to generate the standard curve. The standard curve results are shown in Table 3 and Figure 5. However, in the low-concentration range near the lower limit of quantification (1.5625–3.125 µg/mL), the absolute absorbance values were low, resulting in relatively higher relative error (%RE). This is a characteristic commonly observed near the lower limit of quantification.

Table 3. Standard Curve Results for the 1.5625–25 µg/mL Concentration Range

Input Concentration (µg/mL) Abs. 595 nm Back-calculated Conc.(µg/mL) %RE
1.5625 0.029 1.27 −18.74
3.125 0.063 2.84 −9.21
6.25 0.144 6.57 5.14
12.5 0.285 13.07 4.57
25 0.537 24.69 −1.25

Figure 5
Figure 5. Bradford Protein Assay Standard Curve for the 1.5625–25 µg/mL Concentration Range Using the POP UV-Vis Spectrophotometer A linear regression was applied to generate the standard curve, and the coefficient of determination (R²) was 0.9981.

Conclusion

In this study, standard curves were generated for two concentration ranges, Standard Assay (125–2,000 µg/mL) and Micro Assay (1.5625–25 µg/mL), using Pierce™ Bradford Plus Protein Assay Reagent and the POP UV-Vis spectrophotometer. The results confirm that the POP UV-Vis spectrophotometer can be applied to the Bradford Protein Assay.

References

  • Bradford, M. M. (1976). A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Analytical Biochemistry, 72, 248–254.
  • Thermo Fisher Scientific. Pierce™ Bradford Plus Protein Assay Kit with Dilution-Free™ BSA Protein Standards, Multichannel Pipette Compatible, User Guide. Cat. No. A55866. Thermo Fisher Scientific.
Download PDF