Evaluation of dsDNA Quantification Performance Using the NanoQ Lite+

This technical note evaluated the repeatability and linearity of concentration response of the NanoQ Lite+ in dsDNA mode using dsDNA samples ranging from approximately 3 to 1,500 ng/µL. The measurement results were also compared with those obtained using a commercially available microvolume spectrophotometer. Based on triplicate measurements at each concentration, the %CV was 1.2% or less for samples at concentrations of 11.8 ng/µL and above.
The concentrations measured with the NanoQ Lite+ showed excellent linearity against the dilution-based estimated concentrations, with an R² of 0.9990, and a high correlation with the concentrations measured using the commercially available microvolume spectrophotometer, with an R² of 0.9999. These results demonstrate that the NanoQ Lite+ provides consistent measurement repeatability and a linear concentration response over the tested concentration range, while showing high correlation with the commercially available microvolume spectrophotometer.
Introduction
Nucleic acid bases, the building blocks of DNA, absorb ultraviolet light at approximately 260 nm. At a constant optical path length, absorbance is proportional to sample concentration according to the Beer-Lambert law.
A = εbc
where A is the absorbance, ε is the extinction coefficient, b is the optical path length, and c is the sample concentration. For a dilution series prepared from the same dsDNA sample, the sample composition and effective extinction coefficient remain constant. Therefore, the absorbance at 260 nm is expected to increase proportionally with increasing concentration. Accordingly, the linear change in measured values across a dilution series can be used as an indicator to evaluate the dsDNA concentration response of the instrument.
In general, an absorbance of 1.0 at 260 nm for dsDNA at a 10 mm optical path length corresponds to 50 µg/mL, or 50 ng/µL.[1] In this study, absorbance at 340 nm was used for baseline correction. The dsDNA concentration was calculated by applying a dsDNA conversion factor of 50 to the difference between the absorbance values at 260 nm and 340 nm, normalized to a 10 mm optical path length.
CdsDNA(ng / µL) = (A260 − A340)10mm equivalent × 50
The NanoQ Lite+ is a microvolume spectrophotometer designed for the quantification of nucleic acids and proteins and supports a dsDNA measurement range of 2–1,500 ng/µL. In this technical note, a dilution series of Calf Thymus DNA ranging from approximately 3 to 1,500 ng/µL was used to evaluate the repeatability and linearity of concentration response of the NanoQ Lite+ in dsDNA mode. The initial concentration of the highest-concentration sample used in this study was established based on absorbance measured using a JASCO V-770. In addition, to determine whether the NanoQ Lite+ produced measurement results showing a similar trend to those obtained with a commercially available microvolume spectrophotometer, the same samples were measured using a Thermo Scientific™ NanoDrop™ One for comparison.

Materials and Methods
Sample Preparation
Calf Thymus DNA Solution (Invitrogen, REF 15633019) was diluted 7.5-fold with Nuclease-Free Water (Thermo Scientific, AM9938). The prepared sample was measured in duplicate using a JASCO V-770 UV-Vis/NIR Spectrophotometer with a 1 mm quartz cuvette.
Based on the difference in absorbance between 260 nm and 340 nm, the concentration was calculated by applying a path length correction factor of 10 for the 1 mm cuvette and a dsDNA concentration conversion factor of 50. The resulting mean concentration was 1,507.93 ng/µL, which was used as the initial concentration of the highest-concentration sample in this study.
The highest-concentration sample was then serially diluted two-fold with Nuclease-Free Water to prepare 10 dsDNA samples with estimated concentrations ranging from 2.9 to 1,507.9 ng/µL.
Measurement Procedure
In the dsDNA modes of the K LAB NanoQ Lite+ and Thermo Scientific™ NanoDrop™ One, blank measurements were performed using Nuclease-Free Water. For both instruments, 340 nm was used as the correction wavelength, and the dsDNA concentration was calculated by applying a dsDNA concentration conversion factor of 50 to the difference in absorbance between 260 nm and 340 nm.
Each dsDNA sample (2 µL) was measured in triplicate, in order from the lowest to the highest concentration. The mean, standard deviation (SD), and coefficient of variation (%CV) were calculated from the measurement results. In addition, the linearity of the NanoQ Lite+ measurements against the dilution-based estimated concentrations and their correlation with the measurements obtained using the reference instrument were analyzed.
Results
Repeatability
Using the NanoQ Lite+, dsDNA samples with estimated concentrations ranging from 2.9 to 1,507.9 ng/µL were each measured in triplicate. The mean measured concentration, standard deviation (SD), and coefficient of variation (%CV) at each concentration are shown in Table 1.
Table 1. NanoQ Lite+ dsDNA Repeatability Results
| Estimated Concentration (ng/µL) | Mean Measured Concentration (ng/µL) | SD (ng/µL) | %CV |
|---|---|---|---|
| 2.9 | 3.5 | 0.3 | 7.1 |
| 5.9 | 7.9 | 0.3 | 3.3 |
| 11.8 | 16.6 | 0.2 | 1.2 |
| 23.6 | 32.5 | 0.4 | 1.1 |
| 47.1 | 64.8 | 0.2 | 0.3 |
| 94.2 | 123.6 | 0.4 | 0.4 |
| 188.5 | 226.4 | 0.2 | 0.1 |
| 377.0 | 408.3 | 0.2 | 0.1 |
| 754.0 | 770.9 | 0.8 | 0.1 |
| 1,507.9 | 1,493.5 | 0.8 | 0.1 |
Linear Response to Dilution-Based Estimated Concentration
Linear regression analysis was performed between the dilution-based estimated concentrations and the mean concentrations measured with the NanoQ Lite+. The resulting regression equation was y = 0.9874x + 17.312, with a coefficient of determination (R²) of 0.9990. Here, x represents the dilution based estimated concentration, and y represents the mean concentration measured with the NanoQ Lite+.
With a slope close to 1 and an R² of 0.9990, the NanoQ Lite+ demonstrated a linear measurement response to changes in dsDNA concentration over the range of 2.9–1,507.9 ng/µL.
For some samples in the low- to mid-concentration range, the concentrations measured with the NanoQ Lite+ were higher than the dilution-based estimated concentrations. The same trend was also observed in the measurements obtained with the commercially available microvolume spectrophotometer, as described in the following section. Since the dilution-based estimated concentrations were calculated based on the measured concentration of the highest-concentration sample and the serial dilution factors, they may include uncertainty associated with sample preparation and dilution procedures.
For some samples in the low- to mid-concentration range, the concentrations measured with the NanoQ Lite+ were higher than the dilution-based estimated concentrations. The same trend was also observed in the measurements obtained with the commercially available microvolume spectrophotometer, as described in the following section. Since the dilution-based estimated concentrations were calculated based on the measured concentration of the highest-concentration sample and the serial dilution factors, they may include uncertainty associated with sample preparation and dilution procedures.
Comparison of Measurement Results with a Commercially Available Microvolume Spectrophotometer
To determine whether the measurement results obtained with the NanoQ Lite+ showed a similar trend to those obtained with a commercially available microvolume spectrophotometer, the same dsDNA dilution series was measured using the reference instrument. The mean measured concentrations at each concentration level for both instruments are shown in Table 2.
Table 2. Mean dsDNA Concentrations Measured with the NanoQ Lite+ and Reference Instrument
| Estimated Concentration (ng/µL) | Reference (ng/µL) | NanoQ Lite+ (ng/µL) |
|---|---|---|
| 2.9 | 3.0 | 3.5 |
| 5.9 | 7.2 | 7.9 |
| 11.8 | 15.5 | 16.6 |
| 23.6 | 31.7 | 32.5 |
| 47.1 | 63.4 | 64.8 |
| 94.2 | 121.3 | 123.6 |
| 188.5 | 225.3 | 226.4 |
| 377.0 | 409.8 | 408.3 |
| 754.0 | 748.7 | 770.9 |
| 1,507.9 | 1,476.4 | 1,493.5 |
*Reference instrument: Thermo Scientific™ NanoDrop™ One
For some samples in the low- to mid-concentration range, the concentrations measured with the NanoQ Lite+ were higher than the dilution-based estimated concentrations. The same trend was also observed in the measurements obtained with the commercially available microvolume spectrophotometer, as described in the following section. Since the dilution-based estimated concentrations were calculated based on the measured concentration of the highest-concentration sample and the serial dilution factors, they may include uncertainty associated with sample preparation and dilution procedures.
Conclusion
The NanoQ Lite+ demonstrated consistent repeatability and a linear measurement response to changes in dsDNA concentration over the approximate range of 3–1,500 ng/µL. It also showed a high correlation with a commercially available microvolume spectrophotometer, demonstrating its suitability for dsDNA quantification within the concentration range evaluated in this study.
In addition to its quantitative performance, the NanoQ Lite+ provides simultaneous measurements at A230, A260, and A280 within one second and features a built-in cuvette holder dedicated to OD600 measurements, enabling both nucleic acid quantification and cell culture monitoring with a single instrument.
References
[1] F. H. Stephenson, “Nucleic Acid Quantification,” in Calculations for Molecular Biology and Biotechnology: A Guide to Mathematics in the Laboratory, 2nd ed., Academic Press, 2010, pp. 99–122. https://doi.org/10.1016/B978-0-12-375690-9.00005-X