Relative Quantitation Example Using Indirect ELISA and Application of the MRX A2000 Microplate Reader

This application note demonstrates standard curve generation and data processing using the MRX A2000 microplate reader through a relative quantitation example based on indirect ELISA with the Bio-Rad Giant Panda Problem Kit. A 4-parameter curve fit was generated from serial 2-fold dilutions of primary antibody, yielding R²=0.9977, with the MRX A2000 software providing back-calculated concentrations, %CV, and %RE for each standard solution.
Overview
This application note was prepared to construct a relative quantitation example based on indirect ELISA using the ELISA components of the Bio-Rad Giant Panda Problem Kit, and to verify the standard curve generation and data processing capabilities of the MRX A2000 microplate reader. Although the Giant Panda Problem Kit is originally a qualitative kit intended primarily for visual (naked-eye) readout, in this test the primary antibody was serially diluted and used as standard solutions, and a standard curve was generated based on the resulting change in absorbance.
The input values used in this test, 1.000, 0.500, 0.250, 0.125, 0.062, and 0.031, are not absolute concentrations but rather relative values corresponding to the dilution factor of the primary antibody. Accordingly, this document is not intended as a quantitative ELISA for calculating the absolute concentration of unknown samples; rather, its purpose is to present, as an example, the signal changes according to dilution factor, the application of a 4-parameter curve, and the results calculated by the MRX A2000 software for back-calculated concentration, coefficient of variation (%CV), and relative error (%RE) from the standard curve.

Materials and Methods
Kit and Reagents Used
The Bio-Rad Giant Panda Problem Kit (Cat. No. 17002878EDU) was used for this experiment. The 96-well plate included with the kit was used. The main reagents were as follows.
- 10× PBS
- 10% Tween 20
- HRP Enzyme Substrate (TMB)
- Antigen, Chicken Gamma Globulin, lyophilized
- Primary Antibody, Rabbit Anti-Chicken Polyclonal Antibody, lyophilized
- Secondary Antibody, Goat Anti-Rabbit Antibody HRP, lyophilized
The lyophilized antigen, primary antibody, and secondary antibody were each reconstituted and diluted in 1× PBS to prepare 1× antigen, 1× primary antibody, and 1× secondary antibody solutions.
Buffer and Reagent Preparation
- 1× PBS: Prepared by diluting 10× PBS 10-fold with distilled water.
- Wash buffer: Prepared by adding Tween 20 to 1× PBS to a final concentration of 0.05%.
Standard Solution Preparation
The 1× primary antibody solution was used as the starting solution (STD01), and STD02–STD06 were prepared by serial 2-fold dilution using 1× PBS. The blank consisted of 1× PBS only. The input concentrations in the software were set to 1.000, 0.500, 0.250, 0.125, 0.062, and 0.031, respectively.
Indirect ELISA Procedure
As this kit is designed for rapid educational use, all reaction steps below were performed with shorter reaction times than a conventional ELISA.
- Antigen coating: 50 µL of the 1× antigen solution was dispensed into each well of the 96-well plate used. After incubating for approximately 5 minutes at room temperature, the plate was washed 3 times with wash buffer.
- Primary antibody reaction: 50 µL of the blank and STD01–STD06 solutions were dispensed into each corresponding well. After incubating for approximately 5 minutes at room temperature, the plate was washed 3 times with wash buffer.
- Secondary antibody and TMB reaction: 50 µL of the 1× secondary antibody solution was dispensed into each well and incubated for approximately 5 minutes at room temperature, followed by 3 washes with wash buffer. Then, 50 µL of TMB substrate solution was dispensed into each well, and the plate was measured on the MRX A2000.
Absorbance Measurement and Data Processing
Once color development was complete, the plate was measured on the MRX A2000 in endpoint mode at a wavelength of 655 nm. In the software’s layout editor, the blank and STD01–STD06 were each assigned to duplicate wells, and a 4-parameter curve was applied to generate the standard curve. For each standard solution, the back-calculated concentration, standard deviation, %CV, and %RE were configured to be calculated automatically.
Results
Applying the 4-parameter curve in the MRX A2000 software, the coefficient of determination (R²) was calculated as 0.9977. Table 1 summarizes the input concentration, back-calculated concentration, mean absorbance, %CV, and %RE for STD01–STD06. For STD01–STD05, at input values of 1.000, 0.500, 0.250, 0.125, and 0.062, %CV ranged from 0.603 to 6.104, and %RE ranged from −12.400 to 12.661. For STD06, the calculated concentration for one of the two replicate wells was displayed as “< 0.030”, and as a result, the mean concentration, %CV, and %RE for STD06 were reported as N/A. These results can be seen in the standard curve graph generated by the MRX A2000 software (Figure 3) and the standard data summary screen (Figure 4).
Table 1. Summary results for standard solutions (STD01–STD06)
| Sample Name | Input Concentration | Back-calculated Mean Concentration (Standard Curve) | Mean Absorbance | %CV | %RE |
|---|---|---|---|---|---|
| STD01 | 1.000 | 0.976 | 0.936 | 1.202 | 2.363 |
| STD02 | 0.500 | 0.519 | 0.729 | 6.104 | −3.746 |
| STD03 | 0.250 | 0.237 | 0.365 | 0.603 | 5.308 |
| STD04 | 0.125 | 0.141 | 0.189 | 0.870 | −12.400 |
| STD05 | 0.062 | 0.055 | 0.065 | 6.102 | 1.000 |
| STD06 | 0.031 | N/A | 0.041 | N/A | N/A |
Conclusion
The MRX A2000 microplate reader and its software provided the data processing capabilities needed for standard curve generation and results reporting in an experiment configured as an indirect ELISA, as in this example. This allows the back-calculated concentration, %CV, and %RE for each standard solution to be reviewed on a single screen.
References (Sources)
Bio-Rad Laboratories. Giant Panda Problem Kit for AP Biology: A ThINQ! Investigation. AP Biology Instructor’s Guide. Catalog #17002878EDU. Bio-Rad Laboratories, Hercules, CA, USA.