Example of a Bradford Protein Assay Using the Workflow Feature of the MRX A2000 Microplate Reader

MRX_Example of a Bradford Protein Assay  Using the Workflow Feature of the  MRX A2000 Microplate Reader

In the Bradford Protein Assay, Coomassie Brilliant Blue G-250 binds to proteins and develops a blue color. The absorbance of the resulting protein-dye complex is measured at 595 nm and used to determine the protein concentration of the sample. The Bradford Assay is widely used as a standard method for protein quantification in life science, biochemistry, and molecular biology because it is rapid and simple to perform.
In this application note, standard curves were generated using BSA (Bovine Serum Albumin) standards and Pierce™ Bradford Plus Protein Assay Reagent, and the applicability of the Bradford Protein Assay using the MRX A2000 microplate reader and a 96-well plate was evaluated.
Standard curves were also generated over two concentration ranges, Standard Assay (125–2,000 µg/mL) and Micro Assay (1–25 µg/mL), to confirm the applicability of the assay in each concentration range.

Introduction

The Bradford Protein Assay is a widely used colorimetric method for quantifying protein concentration based on the change in absorbance that occurs when Coomassie Brilliant Blue G-250 dye binds to proteins. The protein-bound dye develops a blue color, and the protein concentration of a sample can be quantified by measuring the absorbance of the resulting complex at 595 nm.

This application note describes the procedure for performing the Bradford Protein Assay using BSA standards, the Pierce™ Bradford Plus Protein Assay Kit, and the MRX A2000 microplate reader. It also demonstrates automated measurement using the MRX A2000 Workflow feature, as well as standard curve generation and data analysis using MRX View 2.0.

Figure 1
Figure 1. K LAB MRX A2000 Microplate Reader

Materials and Methods

Reagents and Instrumentation

The assay was performed using the MRX A2000 microplate reader and MRX View 2.0 software. Pierce™ Bradford Protein Assay Kit (Thermo Fisher Scientific, Cat. No. A55866) was used for protein quantification, and Pierce™ Dilution-Free™ BSA (Bovine Serum Albumin) Protein Standards (REF 1863881) included in the kit were used as standards. Measurements were performed using a 96-well microplate (SPL Life Sciences, Cat. No. 32396).

Experimental Procedure

The Bradford Protein Assay was performed according to the instructions for the Pierce™ Bradford Protein Assay Kit. Standard curves were generated for BSA standards over two concentration ranges: 125–2,000 µg/mL and 1–25 µg/mL. The two concentration ranges were prepared using different volumes of BSA standard solution and Bradford Reagent, as shown in Table 1.

For the 125–2,000 µg/mL range, the Pierce™ Dilution-Free™ BSA Protein Standards included in the kit were used without dilution. For the 1–25 µg/mL range, the standard solution was diluted with distilled water to prepare six concentrations: 1, 5, 10, 15, 20, and 25 µg/mL. Blanks were prepared by dispensing distilled water instead of standard solution at the same volume used for each concentration range.

Table 1. BSA Standard and Bradford Reagent Dispensing Conditions by Concentration Range

BSA Concentration Range (µg/mL) BSA Standard Volume (µL) Bradford Reagent Volume (µL)
125–2,000 5 250
1–25 150 150

The prepared plate was measured using the MRX A2000. Because the color intensity of the Bradford Protein Assay changes with reaction time, applying the same reaction time to all samples is important for ensuring repeatability. Measurements were performed using the Workflow feature. The Workflow feature helps users to combine shaking, delay, measurement, prompt, and other steps in a user-defined sequence to create an automated measurement procedure suitable for the experiment. By executing shaking, reaction, and measurement in a fixed sequence, it eliminates variation in reaction time that may occur during manual operation. In this assay, plate loading, shaking, a 10-minute reaction, and absorbance measurement were configured as a single workflow. An example of the workflow configuration is shown in Figure 2, and the measurement conditions are listed in Table 2.

Figure 2
Figure 2. Example of the MRX A2000 Workflow for the Bradford Protein Assay

Table 2. MRX A2000 Workflow Measurement Conditions

Parameter Condition
Measurement Mode Endpoint
Measurement Wavelength 595 nm
Shaking Linear, Low, 30 sec
Reaction Time 10 min

When the workflow is started, the plate tray opens and a prompt window instructing the user to load the plate is displayed. The user only needs to place the reagent-dispensed plate on the tray and press the confirmation button. The MRX A2000 then performs shaking according to the configured workflow, enables the plate to react for 10 minutes at room temperature, and automatically measures the absorbance.

Figure 3
Figure 3. Automated Bradford Protein Assay measurement using the MRX A2000 Workflow – The user places the reagent-dispensed 96-well plate on the tray and presses the confirmation button. Shaking, reaction, and absorbance measurement are then performed automatically according to the configured workflow.

Results

Standard curves were generated using BSA standards over two concentration ranges: 125–2,000 µg/mL and 1–25 µg/mL. Mean absorbance values were calculated after subtracting the mean absorbance of the blank. Standard curve generation and data analysis were performed using MRX View 2.0.

For the 125–2,000 µg/mL concentration range, each concentration was measured in triplicate (n=3). The results showed that the increase in absorbance gradually became less pronounced as the concentration increased in the higher concentration range. Therefore, rather than interpreting the entire concentration range using a single linear regression equation, a second-order polynomial regression was applied to generate the standard curve. The color development results are shown in Figure 4, and the standard curve results are shown in Table 3 and Figure 5.

Figure 4
Figure 4. Bradford Protein Assay color development over the 125–2,000 µg/mL concentration range – The blue color became more intense as the protein concentration increased.

Table 3. Standard Curve Results for the 125–2,000 µg/mL Concentration Range

Input Concentration (µg/mL) Mean Absorbance Mean Calculated Concentration (µg/mL) Calculated Concentration (%CV)
125 0.0793 127.71 3.38
250 0.1670 249.26 2.79
500 0.3422 510.76 2.09
750 0.4803 739.35 1.83
1,000 0.6037 966.64 3.10
1,500 0.8624 1,564.63 2.52
2,000 0.9822 1,967.13 3.07

Figure 5
Figure 5. Bradford Protein Assay standard curve for the 125–2,000 µg/mL concentration range generated in MRX View 2.0 – A second-order polynomial regression was applied to generate the standard curve, with a coefficient of determination (R²) of 0.9986.

For the 1–25 µg/mL concentration range, each concentration was measured in duplicate (n=2). Absorbance and concentration showed a linear relationship within the measurement range. Therefore, linear regression was applied to generate the standard curve. The color development results are shown in Figure 6, and the standard curve results are shown in Table 4 and Figure 7.

Figure 6
Figure 6. Bradford Protein Assay color development over the 1–25 µg/mL concentration range

Table 4. Standard Curve Results for the 1–25 µg/mL Concentration Range

Input Concentration (µg/mL) Mean Absorbance Mean Calculated Concentration (µg/mL) Calculated Concentration (%CV)
1 0.0205 1.09 17.60
5 0.0773 4.29 12.86
10 0.1843 10.33 1.03
15 0.2770 15.56 0.82
20 0.3612 20.31 1.22
25 0.4340 24.42 0.52

Figure 7
Figure 7. Bradford Protein Assay standard curve for the 1–25 µg/mL concentration range generated in MRX View 2.0 – Linear regression was applied to generate the standard curve, with a coefficient of determination (R²) of 0.9967.

Conclusion

In this study, the Bradford Protein Assay was performed using the Pierce™ Bradford Plus Protein Assay Kit and the MRX A2000, and standard curves were generated by applying regression models appropriate for each concentration range. The results also confirmed that the Bradford Protein Assay can be performed efficiently through standard curve generation and data analysis using MRX View 2.0 and automated measurement using the MRX A2000 Workflow feature.

References (Sources)

  • Bradford, M. M. (1976). A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Analytical Biochemistry, 72, 248–254.
  • Thermo Fisher Scientific. Pierce™ Bradford Plus Protein Assay Kit with Dilution-Free™ BSA Protein Standards, Multichannel Pipette Compatible, User Guide. Cat. No. A55866. Thermo Fisher Scientific.